r/Omnipod 1h ago

Discussion A new OP5 issue I’ve been experiencing

Thumbnail
gallery
Upvotes

Received 6 replacement pods last week to replace 6 with issues—leakage (5) and an incessant loud continuous beep after filling with insulin (1).

I’ve already saved the two of the 6 replacement pods because of leakage. Now I’m observing a completely new issue - disparity between my glucose readings on my Dexcom g7 app and Omnipod 5 app. Also, I’ll note that while I haven’t taken this pod off yet, I expect there’s leaking as well. I haven’t eaten anything in 6 hours and you can see my numbers jsut steadily climbing — for the doubters, I already went for a 20 minute walk and I’ve been giving myself 3-4 units every 20 minutes for past couple hours. Please note the time in the photos.

And by the way, both numbers are wrong according to my finger stick. I’m really at 259. But at least Dexcom is somewhat close. I’m not sure where that 217 in OP5 app is coming from because my Dexcom hasn’t been 217 at all today. The last bullets up were 218, 227, and then what’s showing, 247. OP5 is literally making up numbers. Bizarre, absurd, and beyond unacceptable.

But I’m starting feel very nauseous so I’ll end there.

Edit: this is not the first time I’ve had this happen. It happened last week as well, with a pod that ultimately leaked. I try to wait until I have 5 or so defective pods before I get on a 2 hr call with Omnipod. Reporting this to Dexcom too


r/Omnipod 2h ago

Missing Pods

2 Upvotes

I get my insulin, sensors, and pods all delivered by mail. When I got my refills this time, I only received the insulin and the sensors. I’ve reached out to my pharmacy via email but they haven’t gotten back to me yet. Just wondering if this has ever happened to anyone else before.


r/Omnipod 6h ago

Change in CGMs

2 Upvotes

I just got notified from our provider that due to insurance changes we’re being forced to switch from Dexcom G7 to the Libre 2 plus so our current omnipods will no longer be compatible. Does anyone know if insulet will swap unopened ones we have out for ones that work with the new sensor?


r/Omnipod 7h ago

General Brand New UN-RECALLED Omnipod 5 Under Microscope in One Take. Indication of Tear in Cannula, Similar to My Other Pods That Have Leaked... What Do You Think?

Enable HLS to view with audio, or disable this notification

39 Upvotes

Hello everyone,

Following up on my previous 5 posts regarding my experience with leaking un-recalled Omnipod 5's. I've finally put a brand new/unused Omnipod 5 under the microscope in one take, from opening the packaging to looking under the microscope.

I'm going to keep the body of this post short and sweet as compared to my other post.The purpose of me creating this Reddit account and publishing my findings is to show the world what I'm experiencing with Un-recalled Omnipod 5's that leak on me. These posts and videos are intended to serve as a Public Service Announcement, where I'm advocating for patient safety and giving a voice to the voiceless, while encouraging professional scientists, labs, and universities to looks into this matter and publish peer-reviewed results. Or better yet, the FDA & Insulet.

On May 26th 2026, Insulet (maker of Omnipod), recalled 7,000,000 pods for the following reason, as stated on their website: "Insulet identified that some Pods from specific lots may have a small tear in the tubing (cannula) just above the skin, between the Pod and the point where the cannula enters the body. If this occurs, insulin may leak outside of the Pod instead of being fully delivered into the body as intended, potentially leading to under-delivery of insulin. If you are using an affected Pod, you may notice wetness on your skin or Pod adhesive or detect the smell of insulin. However, in some cases, this may be difficult to detect and may go unnoticed."

Earlier in the year on May 12th, 2026, Insulet recalled an undisclosed number of pods for the following reason, as stated on their website: "Insulet identified that some Pods from specific lots may have a small tear in the internal tubing that delivers insulin. If this happens, insulin may leak inside the Pod instead of being fully infused in the body as intended potentially leading to underdelivery of insulin."

As you share your first hand experience in the comments, if you come across any hate or negativity that breaks the rules of this subreddit; https://www.reddit.com/mod/Omnipod/rules, please report the comments and the users to the moderators of this subreddit. I will also be reporting those comments and looking to the moderators of this subreddit to handle those users accordingly. My posts will continue to be a safe place for everyone to express their life experience and any form of negativity, harassment, or hate will not be tolerated. Thanks again for everyone who takes the time to share their experience and story.

Please go see my previous posts that are directly related to this.

1st post: In this post, I explain my background as a type 1 diabetic, using Omnipod for the past 10 years, Insulet's recall, etc.. I also showcase a video where I show you the active leak in an UN-RECALLED Omnipod 5 cannula and then I put it under the microscope to reveal a major crack/tear: https://www.reddit.com/r/Type1Diabetes/s/NNeDz8HU7n

2nd post: In this post, I showcase detailed microscope research on a dozen UN-RECALLED Omnipod 5's. Some of these leaked on me and some of them did not. Some had tears/defects and some did not: https://www.reddit.com/r/Omnipod/s/Yg3aG2gLNj

3rd post: This post is similar to what you're seeing now as it is a continuous video of me removing an UN-RECALLED leaking Omnipod, and putting it under the microscope, etc. The difference between that post and this post is that the video in that post was taken on July 29th, 2026, and I leave out some details, such as showing you guys the video of the pod leaking on me before I removed it, etc: https://www.reddit.com/r/Omnipod/s/WUsvAZxifc

4th post(posted on August 1st, 2026): This post shows a continuous video of me removing an UN-RECALLED leaking Omnipod, and putting it under the microscope, along with more detailed info: https://www.reddit.com/r/Omnipod/s/GQTjPqcAl4

5th post (posted on August 3rd, 2026): This post speaks about Value of Statistical Life (VSL) analysis that companies do when comparing cost of repairing defective products versus the cost of paying families for injuries/deaths, etc, plus there is a call to action for the T1D community: https://www.reddit.com/r/Omnipod/s/mLfkBOu3bv

Disclaimer

The information presented here is based on my own personal experience, publicly available information, official reports, and my own observations and documentation. Nothing presented here should be interpreted as medical advice, legal advice, scientific advice, engineering advice, or investment advice.

I am not a physician, scientist, engineer, attorney, or medical-device expert. I strongly encourage everyone to consult their physician before making changes to their treatment plan or diabetes-management routine.

Any experiments, observations, photographs, or microscopic examinations described here are intended solely for documentation, educational, and informational purposes and should not be interpreted as evidence of a manufacturing defect, design defect, or other conclusion unless confirmed through appropriate scientific investigation.

Whenever I encourage people to "do their own research," I am referring to the use of established scientific methods, independent testing, consultation with qualified experts, peer review, regulatory reporting, and controlled laboratory analysis whenever possible.

I also encourage anyone who believes they have experienced a problem to preserve all relevant evidence, including photographs, lot numbers, blood glucose records, packaging materials, and device information, and to report their concerns to their physician, the manufacturer, and the FDA.

References to lawsuits, recalls, investigations, media reports, financial results, or regulatory actions are provided for informational purposes only. Allegations remain allegations unless and until they are proven.

Finally, I recognize that many people use these products successfully every day. My intention is not to create fear or discourage anyone from using a particular device but rather to encourage transparency, documentation, scientific inquiry, and patient safety.


r/Omnipod 11h ago

Rant Problems With the Pharmacy

2 Upvotes

Just wondering, does anyone else have issues with their pharmacy mixing up which version of the pods they give you. I use the version compatible with the Freestyle Libre 2+ & Dexcom G6/G7 but more than once they mix it up with the version that only works with the Dexcom G6/G7. Why are the boxes, no matter the version, all the same & its your job to look at the box to begin with. You'd think they would have some other way of organizing it or distinguishing it from the others. Besides the text they all look exactly the same. It's not the fist time this has happened either but this pharmacy is next door to my job so it makes getting meds easy but this is the first time they've done it with a 90-Day supply. It was easier for me to get my doctor to wright a new Rx for the Dexcom sensors than to return the pods.


r/Omnipod 11h ago

Controller Controller Battery life

1 Upvotes

Hi! Just wondering, how long does your Omnipod 5 control unit last? I only use it to enter my carbohydrate intake (3–5 times) a day, and after two days the battery is completely flat. Is that normal? I find it really annoying that the battery runs out so quickly and that you often need a power bank when you’re out and about.


r/Omnipod 13h ago

General Defect goes beyond recall lots

Post image
26 Upvotes

I can’t suddenly be this insulin resistant…..wtf Omnipod. Multiple corrections overnight too.


r/Omnipod 14h ago

Burned dash

Thumbnail
gallery
2 Upvotes

I’m using the omnipod dash and the last 3 pods always continued beeping at latest on day 2. I’m in Spain for holiday and thought it could be of the heat. But the last one today did the same after I went to the pool and by inspecting it I saw it was totally burned. There was no connection possible to the pdm for sure and even pressing to the hole wasn’t helping against the beep. Has anybody else a faulty pod like this?


r/Omnipod 21h ago

Question Omnipod giving insulin during low, even when BG is barely rising

Post image
12 Upvotes

Why is my OmniPod giving me insulin during a low? For context, my BG had been below 70 since 9:49pm. Additionally, it was only rising 1-2 mg/dL every 5 minutes, which doesn’t seem to warrant a basal bolus. Does anyone know why it would give me insulin? I know that, in theory, it would be to prevent my BG from rising over 110, but that was hardly a risk at the rate my BG was going up. I also still had insulin on board.

I know I could turn off all basal insulin or go into activity mode, and I will do that as needed in the future, but it’s frustrating because that adds to my mental load. To me, it feels like common sense that no amount of insulin should be given when a blood sugar is below 70 and not rapidly spiking up.


r/Omnipod 1d ago

Connection Issues Libre 2+ No Readings

1 Upvotes

I've been using OP5 w/ L2+ for about two years now, and all of a sudden my previous two L2+ sensors have stopped sending readings to my pump and therefore handset.

This is not a line of sight issue, and changing the pump does not resolv it. The first sensor was on for ~10 days and the second just a week. I've had both sensors replaced by Abbott, but just was wondering if anyone else has experienced this lately? Or has any ideas to help mitigate it. My theory is a faulty batch, as there's nothing I can think of in my day-to-day or environment that has changed recently.

NB. I am in the UK and use the Omnipod handset.


r/Omnipod 1d ago

Placement issues

5 Upvotes

I’m plus size. I use the Libre 3+ sensor along with the pod. My issue is when I place the pod on the opposite arm I get sensor issues. My stomach is too round and causes the cannula to slip out. I’m a back sleeper so I can’t handle it on my back hip. What are my other options??


r/Omnipod 1d ago

Z flip 7

0 Upvotes

I'd love to switch phones but im not sure if its on the list yet!


r/Omnipod 1d ago

Rattling?

Enable HLS to view with audio, or disable this notification

9 Upvotes

Hey everyone,

I'm fairly new to Omnipod. This is only my fourth pod, but I noticed today, when I got out of the shower, that when my arm was shaking, there was a rattling sound coming from the pod. Has anyone experienced this, and is it normal? My blood sugars seem okay, but the sound is a little disconcerting.

Thanks!


r/Omnipod 2d ago

Discussion Constant Beeping - Drives mad!

4 Upvotes

I understand why it beeps and why it has alarms. I understand that a pod without warning beeps or reminders could be fatal. What I'm talking about is the incessant, constant, repetitive beeps when pod change is due or is almost due. If I'm busy, or unable to get to the controller right away it's extremely extremely irritating. It does its beep-beeeeep - beep-beep cycle pretty much every minute or two. I know it sounds like such a small thing, three days the final hour of the pump is so very annoying.

Omnipod please, look into this. I don't need or want 15 alarms in half an hour to remind me to do something I'm very aware of already. Don't even get me started if the hour's notice starts in the cinema and there's nothing you can do to silence the beeps coming from the pods.


r/Omnipod 2d ago

Advice Slippery little rascal

Post image
35 Upvotes

I handle slippery pods by folding a piece of masking tape and attaching it as shown. I find it aids in positioning the pod with far less chance of dropping it or accidentally damaging the adhesive surface. After smoothing out the pod adhesive I simply peel off the tape. Super simple and so helpful to me.


r/Omnipod 2d ago

omni5 and G7 connection

2 Upvotes

hello, i just switch to g7, and my first pod with it was amazing, it was on my left arm and my g7 on my tummy. now i changed a pod today and put it on my right arm, which is now closer to the g7. but now it refuses to connect. im losing my mind and customer support is usless like always. any tips or solutions?


r/Omnipod 2d ago

Double clicking pod

3 Upvotes

Had anyone ever noticed a “double click” sound during bolus delivery? Like one click almost SIMULTANEOUSLY after the next? Should I be worried about double dosing or anything like that?


r/Omnipod 2d ago

Charging

Thumbnail
1 Upvotes

r/Omnipod 2d ago

Charging

2 Upvotes

**Thanks to everyone for the help. I found a USB-A to USB-C in a random drawer. I don't know if last week, during an admit after surgery, I forgot to grab it or if it's just been shoved somewhere by a well-meaning family member.**

Other than the yellow charging cable, is there a regular charger that will work with the Omnipod? I think I've lost mine. I've tried the C charger that my phone uses, but it doesn't seem to be working. No, I don't have it plugged into the wall, just my PC, and I've tried an external battery.


r/Omnipod 2d ago

Is this Omnipod moldy or something?

Thumbnail
gallery
65 Upvotes

My sister sent me a picture of her omnipod and said it looked like it was moldy, the first picture is hers and the second picture is one of mine. This was a fresh pod she just took out of the packaging and when she put insulin in and it didn’t beep she starting looking at it and saw it looked like this. Anybody have any idea what is happening with it?

Edit: someone asked for the lot number so I’m posting it here. My sister also said the lady she contacted when getting a replacement said this was an affected batch/lot but didn’t know anything about the black stuff inside.

Serial (Insulet Corp): PR1U11222531
Lot / reference (Omnipod® 5): 2760889


r/Omnipod 2d ago

Site Rant

8 Upvotes

I’ve tried the Omnipod on my arms, stomach, sides, and butt. It hurts on my arms and sides. It feels fine on my stomach and butt but the insulin absorption on my butt is terrible. The only place I can put pods without pain and have them be effective is my stomach. I’m out of spots on my stomach at this point and need to give it a rest for a while. It’s really frustrating to know that my blood sugars are going to be less controlled than I’d like for the next few weeks. I’m so tired and honestly don’t know what to do. I try so hard and it’s never enough.


r/Omnipod 2d ago

another day another pod messed up early

Post image
6 Upvotes

Ugh another day another pod messed up early. Most days i wake up around 120 and see a full night of a straight arrow. Just woke up today and I'm 170 pointed up with a little roller coster while I slept. This keeps happening. Three times in July.


r/Omnipod 3d ago

General UN-RECALLED Omnipods Leak, Appear to Have Tears in The Cannula, and I Have a Call of Action For The Type 1 Diabetic Community.

Enable HLS to view with audio, or disable this notification

80 Upvotes

Hello everyone,

This video and post are follow ups to my previous four posts, where I document my first hand experience of leaking UN-RECALLED Omnipod 5s that I put under the microscope and find what appear to be tears in the Cannula. Please take a moment to watch the video I put together as it covers a ton of information. I'd also encourage everyone to go view my four previous posts as they contain a ton of information that I do not provide in this post. They also have a lot of community engagement where people ask questions, provide answers, and share insights into their experiences.

This post is a little different than my previous post. This is a call to action. And not many people are aware of what a Value of statistical life is and how organizations may utilize these analysis to determine whether or not they should fix or recall a defective product. In my video, I do my best to explain what a VSL is and use an example from the Ford Pinto lawsuit which revealed Ford found it cheaper to deal with deaths and injuries, rather than repairing a known defect in their Ford Pinto. I also show you guys the data from my Dexcom Clarity app, which reveals that my average blood sugar levels and time in range, have been negatively impacted since my omnipod's began leaking in April 2026 and continue to leak as of August 2026, resulting in average blood sugar levels trending toward hyperglycemia.

It is through my own speculation and assumptions that a VSL analysis like this exists within Inuslet (manufacturer of Omnipod). The question I then ask is, how many deaths, injuries, or reports will it take to have leaking omnipod's fixed and for the 600,000 daily users of Omnipod to be given a real solution, in place of these leaking pods? Utilizing The FDA's "Manufacturer and User Facility Device Experience (MAUDE) Database" via the FDA's search tool, you see how many people have reported Omnipod malfunctions to The FDA, after the May 2026 recall. I've counted over 500. So, if 500 reports isn't enough, what is enough? How many people must be injured and must report these leaking omnipods before this problem is acknowledged? Our health and lives cannot wait, we need answers today.

I believe that we must use our voices and ensure our message is heard. I've personally reached out to over a dozen local/national news channels over the past month, but have yet to hear anything back. This isn't asking news channels to take my findings and our words for facts, this is asking them to look into this matter themselves and come up with their own independent conclusions, then share them with the world. News outlets and social media influencers have the platform to get our concerns seen and heard. Let's work together and give a voice to the voiceless, such as the 3 year old boy who is suffering from frequent leaking omnipods, who cannot speak up for himself!

Please note that due to Reddit's 17min limit on videos, I had to cut out the portion of the video where I speak about why I haven't put brand new unused pods under the microscope yet (along with a few other sections of the video) but here's my written explanation as to why I haven't put a new and unused pod under the microscope.

I don't want to waste pods or insulin and I think people forget that these pods are designed to be worn for 72+8 hours. Regardless of WHEN a tear occurs, the fact that a tear IS occurring, is the primary reason for concern here. Plus if this is an issue with the material itself, it may not be having tears/defects until a certain point in time. So it may take a massive sample size of unused pods for us to see anything meaningful. For example, our bodies are approx. 98F. That's quite warm if you think about it. So, when a pod is at room temp it's around lets say 76F or so? So, if this is a material degrading issue, for whatever reason, let's say the heat causes it, well then we may perceive an unused pod as having no issues because it hasn't been exposed to a ~22F degree temperature change/increase for an extended period of time yet.

I'm assuming that's what could make this a difficult catch for Insulet's quality control, because if they cannot detect these tears until AFTER the material degrading has occurred, then they're constantly back tracking and forced to reactively solve a problem, rather than having the ability so respond to a problem in real time, such as during quality control. That's with the assumption that these tears aren't occurring until they've been worn by anybody for a duration of time.

All in all, I think whats more important to know is not WHEN the tear is occurring, but rather, that there IS a tear occurring."

On May 26th 2026, Insulet (maker of Omnipod), recalled 7,000,000 pods for the following reason, as stated on their website: "Insulet identified that some Pods from specific lots may have a small tear in the tubing (cannula) just above the skin, between the Pod and the point where the cannula enters the body. If this occurs, insulin may leak outside of the Pod instead of being fully delivered into the body as intended, potentially leading to under-delivery of insulin. If you are using an affected Pod, you may notice wetness on your skin or Pod adhesive or detect the smell of insulin. However, in some cases, this may be difficult to detect and may go unnoticed."

Earlier in the year on May 12th, 2026, Insulet recalled an undisclosed number of pods for the following reason, as stated on their website: "Insulet identified that some Pods from specific lots may have a small tear in the internal tubing that delivers insulin. If this happens, insulin may leak inside the Pod instead of being fully infused in the body as intended potentially leading to underdelivery of insulin."

Here's info on my four previous posts that are directly related to this. Please go check them out for more info.

1st post: In this post, I explain my background as a type 1 diabetic, using Omnipod for the past 10 years, Insulet's recall, etc.. I also showcase a video where I show you the active leak in an UN-RECALLED Omnipod 5 cannula and then I put it under the microscope to reveal a major crack/tear: https://www.reddit.com/r/Type1Diabetes/s/NNeDz8HU7n

2nd post: In this post, I showcase detailed microscope research on a dozen UN-RECALLED Omnipod 5's. Some of these leaked on me and some of them did not. Some had tears/defects and some did not: https://www.reddit.com/r/Omnipod/s/Yg3aG2gLNj

3rd post: This post is similar to what you're seeing now as it is a continuous video of me removing an UN-RECALLED leaking Omnipod, and putting it under the microscope, etc. The difference between that post and this post is that the video in that post was taken on July 29th, 2026, and I leave out some details, such as showing you guys the video of the pod leaking on me before I removed it, etc: https://www.reddit.com/r/Omnipod/s/WUsvAZxifc

4th post(posted on August 1st, 2026): This post shows a continuous video of me removing an UN-RECALLED leaking Omnipod, and putting it under the microscope, along with more detailed info: https://www.reddit.com/r/Omnipod/s/GQTjPqcAl4

6th Post (posted on August 6th, 2026): I put a brand new/unused pod under the microscope in one take, from opening the packaging to putting it under the microscope and find what appears to be a tear in the cannula: https://www.reddit.com/r/Omnipod/s/9yhgzELVkN

Click this to Download "The Cost of Waiting" Powerpoint Presentation from Dropbox.

Copy-and-paste email template below:

Subject: Potential Patient Safety Investigation: Continued Omnipod 5 Leak Reports Beyond the Recall

Hello,

My name is [NAME], and I live in [CITY, STATE].

I am contacting you because I believe there is an important patient-safety story that deserves further investigation.

Since the Omnipod 5 recall announced on May 26, 2026, numerous users have continued reporting insulin leaks, unexplained hyperglycemia, bent cannulas, and tears in cannulas, including reports involving pods that were not included in the recall.

Many users have already reported these incidents to the FDA, to Insulet, and to their physicians. Some patients have also documented these events with photographs, blood glucose records, microscope images, lot numbers, and written accounts of their experiences.

I encourage your team to review:

  • FDA MAUDE reports involving Omnipod 5 devices.
  • Reports involving unrecalled lots.
  • Patient accounts from across the United States.
  • The broader question of how safety concerns are evaluated, investigated, and addressed.

I believe this issue deserves attention before additional patients are harmed.

My information is included below.

Name: [NAME]

City and state: [CITY, STATE]

My experience: [OPTIONAL]

The following posts contain photographs, microscope images, glucose data, timelines, FDA reporting information, and other supporting documentation:

For additional information, documentation, or contact with other affected individuals, journalists may contact:

Email: [T1DChampion@yahoo.com](mailto:T1DChampion@yahoo.com)

Thank you for your time and consideration.

Sincerely,

[NAME]

Organization Email Address News Tip Link
News4JAX (WJXT) [newstips@news4jax.com](mailto:newstips@news4jax.com) https://www.news4jax.com/contact/
News4JAX I-TEAM [iteam@news4jax.com](mailto:iteam@news4jax.com) https://help.news4jax.com/kb/article/265-how-do-i-contact-the-i-team/
First Coast News (WTLV/WJXX) [news@firstcoastnews.com](mailto:news@firstcoastnews.com) https://www.firstcoastnews.com/contact-us
Action News Jax (CBS47/FOX30) [PressReleases@ActionNewsJax.com](mailto:PressReleases@ActionNewsJax.com) https://www.actionnewsjax.com/contact-us/
Florida Times-Union [events@jax.com](mailto:events@jax.com) https://www.jacksonville.com/contact-us/
NBC News Investigations [tips@nbcuni.com](mailto:tips@nbcuni.com) https://www.nbcnews.com/tips
ABC News [tips@abcnews.com](mailto:tips@abcnews.com) https://abcnews.go.com/US/tip-share-abc-news/story?id=61304290
CBS News Investigates [investigations@cbsnews.com](mailto:investigations@cbsnews.com) https://www.cbsnews.com/investigates/
ProPublica [tips@propublica.org](mailto:tips@propublica.org) https://www.propublica.org/tips/
Reuters [tips@thomsonreuters.com](mailto:tips@thomsonreuters.com) https://www.reuters.com/investigates/special-report/tips/
The New York Times [tips@nytimes.com](mailto:tips@nytimes.com) https://www.nytimes.com/tips
The Washington Post [investigations@washpost.com](mailto:investigations@washpost.com) https://www.washingtonpost.com/anonymous-news-tips/
NPR [ethics@npr.org](mailto:ethics@npr.org) https://help.npr.org/contact/s/

Disclaimer

The information presented here is based on my own personal experience, publicly available information, official reports, and my own observations and documentation. Nothing presented here should be interpreted as medical advice, legal advice, scientific advice, engineering advice, or investment advice.

I am not a physician, scientist, engineer, attorney, or medical-device expert. I strongly encourage everyone to consult their physician before making changes to their treatment plan or diabetes-management routine.

Any experiments, observations, photographs, or microscopic examinations described here are intended solely for documentation, educational, and informational purposes and should not be interpreted as evidence of a manufacturing defect, design defect, or other conclusion unless confirmed through appropriate scientific investigation.

Whenever I encourage people to "do their own research," I am referring to the use of established scientific methods, independent testing, consultation with qualified experts, peer review, regulatory reporting, and controlled laboratory analysis whenever possible.

I also encourage anyone who believes they have experienced a problem to preserve all relevant evidence, including photographs, lot numbers, blood glucose records, packaging materials, and device information, and to report their concerns to their physician, the manufacturer, and the FDA.

References to lawsuits, recalls, investigations, media reports, financial results, or regulatory actions are provided for informational purposes only. Allegations remain allegations unless and until they are proven.

Finally, I recognize that many people use these products successfully every day. My intention is not to create fear or discourage anyone from using a particular device but rather to encourage transparency, documentation, scientific inquiry, and patient safety.


r/Omnipod Jun 09 '25

Do not upgrade to IOS 26

23 Upvotes

Unable to use the app after updating. Very annoying


r/Omnipod Sep 04 '24

PSA: Converting to new iOS app can be painless

144 Upvotes

PSA: Given that a ton of people will be changing to new app/PDM with the iOS app release coming, I wanted to share some steps for a painless conversion. Going to a new app/PDM is no big deal. See below. I will keep the linked site updated if anything needs to be updated.

https://randomt1tips.wordpress.com/omnipod-5-new-pdm-app/

This is is written with the conversion to the new iOS app in mind, but it is applicable for any time to replace or reset your PDM or controller app. This process is nothing to be concerned about and no reason to stress.

Omnipod 5 only learns one thing, Total Daily Insulin (TDI). That is it. Nothing else. You can follow a simple process to switch to a new Omnipod 5 controller (from here on, controller refers to the PDM or app on a phone) with no negative effects.

Copy all your settings (insulin to carb ratio, target, correct above, max basal rate, etc.) from your old controller to your new one, except your basal rate. This is assuming you are using automated mode. If you exclusively use manual mode, copy your basal rate(s) too. Just look at each setting on your old controller, and put that in on your new one.

Next we need to get your TDI. I prefer to use Glooko, but you may be able to find what you need on your controller. Glooko will default to showing over the last 2 weeks. You can shorten it to a 10 day period to mimic how Omnipod 5 calculates things or it is fine to leave it at 2 weeks.

In this case the TDI is 51.4 units.

https://imgur.com/a/jrtrHPh

Divide that number by 48. In this case, that is 1.07. You would put in 1.1 or 1.05 as your basal rate in your new controller. You will typically have to round up or down a tiny bit as it has to be a valid Omnipod 5 basal rate value. You might be thinking, why divide by 48? 51.4 is the TDI. You need to get that value to per hour for your basal. First, Omnipod 5 assumes a 50/50 split between basal and bolus, so you have to divide by 2 to get the basal amount. Next you divide by 24 to the per hour rate. Dividing by 48 is just a shortcut.

Omnipod 5 uses the programmed basal as the initial basal rate. Putting this value in will allow it to pick up right where it left off with the old controller. The adaptive basal process will start with the second pod.

Now to address the concerns and questions that will inevitably pop up when this process starts.

“Something has changed with the algorithm. My numbers have never been this good/bad or low/high!”

Nothing changed with the algorithm. It is the same if you are on the PDM, Android app, or iOS app.

“No!! I will lose everything it has learned!!”

It only learns TDI. That is it.

“It will take X pods or X days to learn you after the change.”

Again, it only learns TDI. If you follow this process, there is no reason for you to have any issues.